K-wire fixation is indicated for:
K-wire fixation alone is not indicated in:
K-wire size is chosen according to the age of the child and the size of the fragment.
The entry point, together with the correct direction of the K-wire, is the key to optimal fixation.
For most simple fractures, two, occasionally three, K-wires give sufficient stabilization if the K-wires:
K-wire osteosynthesis usually requires additional plaster cast protection.
Advantages:
Disadvantages
The following points influence the size of the K-wire:
In children younger than 5–6 years, 1.6 mm K-wires are used for fractures around the shoulder, elbow, knee, and ankle joints.
In children above this age, 2.0 mm K-wires are usually used.
It is important to consider the weight of the patient when choosing the diameter of the K-wire.
Metaphyseal fractures of the long bones require at least 1.6 mm K-wires. If only two wires are used, larger diameter wires may be necessary.
Fractures of small bones (hand and foot) require 1.0–1.6 mm K-wires.
The size of the K-wire should be chosen according to the size of the fragment. For example, a fracture of the medial epicondyle of the humerus requires a K-wire of smaller diameter than a fracture of the lateral humeral condyle.
For fractures fixed with two (or three) K-wires from only one side, one size larger K-wires are used than for bilateral crossed K-wiring.
For example, for bilateral crossed K-wiring of a supracondylar humeral fracture, 1.6 mm K-wires can be used, whereas for radial divergent wiring of the same fracture, 2.0 mm K-wires are preferable.
K-wires are, in most cases, inserted from the free fragment into the main fragment. This allows the K-wire to be used as a joystick for manipulating the free fragment.
The entry points of the K-wires should be chosen so that they are as far apart as possible where they cross the fracture line.
This guarantees maximal rotational stability.
The choice of the entry point must correlate with the planned direction of the K-wire and the end fixation point in the main fragment.
Ideally, if the anatomical site permits, the K-wires should be introduced as perpendicular as possible to the fracture plane. In certain sites, this is not achievable and mechanical stability should not be compromised by obsessive adherence to the above principle.
The direction of the K-wires should be chosen so that the K-wires are well separated at the fracture level.
To achieve this, the length of the fracture line is divided into four equal parts. Ideally, when using two K-wires, the wires should pass approximately through the green areas in the illustration.
For oblique (>30°) metaphyseal fractures crossed K-wire fixation may be very difficult, or impossible, as at least one of the K-wires will run nearly parallel to the fracture line.
For oblique fractures, therefore, divergent monolateral K-wire fixation is more suitable. For this technique, one size larger K-wires should be used than for cross K-wiring.
If lateral divergent K-wire fixation is not possible, for example, due to soft tissue condition or a structure at risk, another stabilization technique should be used (eg, external fixator or plate).
A small incision or a direct puncture with the K-wire is made over the planned entry point. An incision is recommended to avoid skin damage, which might cause pin-track infection.
To avoid thermal injury, especially to the physis, K-wires should be inserted by hand or using an oscillating drill.
If a standard drill is used, it must be run as slowly as possible to avoid a thermal effect.
Additionally, irrigate the K-wire during drilling with a cooled irrigation fluid.
If a drill is used, the K-wire is initially inserted manually through the skin incision, onto the chosen bony entry point. While maintaining the correct position of the tip, the drill is attached to the wire.
To prevent bending of the K-wire, it can be helpful to insert the K-wire using an appropriate drill sleeve – this steadies the wire, protects the soft tissues, and ensures optimal direction.
It is helpful to reduce the length of the K-wire protruding from the drill to avoid whipping of the wire and loss of trajectory.
To avoid skidding, the K-wire tip should initially be held as orthogonal as possible to the bone surface until the tip of the wire has a good purchase.
Once the tip of the K-wire has obtained a good purchase, the angulation of the K-wire should be corrected according to the planned direction of the K-wire.
As soon as increased resistance is felt, check that the tip of the K-wire is engaged in the far cortex of the main fragment.
The tip of the K-wire should penetrate the whole depth of the far cortex, but not protrude more than 2–3 mm. This is to avoid neurovascular damage and soft tissue irritation.
The free end of the wire is usually left protruding through the skin and is bent through 180°. A sterile dressing protects the entry wound around the wire.
No more than two attempts should be made to insert any one wire across a physis. Repeated puncture of the physis by multiple attempts to insert the wire can result in subsequent growth disturbance.
The timing of K-wire removal is a matter of judgment by the treating surgeon, based on the age of the child, the pattern of the injury, as well as additional injuries.
Depending on the age of the child, fracture healing has reached the stage where redisplacement is highly unlikely after 3–4 weeks and the K-wires can be removed.